Reasons for incorrect display of vortex flowmeter

Reasons for incorrect display of vortex flowmeter

         The reasons for the inaccurate display of vortex flowmeters are mainly caused by inaccurate parameters, selection and field installation problems. Users of flowmeters should read the manual before installing the meter.

、如何选型: 1. How to choose:

The selection of vortex flowmeter is mainly based on the parameters provided by the user to select the appropriate instrument, and the selection is based on the five parameters of medium, flow, caliber, pressure and temperature.

、安装问题 2 , installation problems

The installation of the vortex flowmeter must ensure the straight pipe section before and after. Do not install some accessories such as valves and joints that affect the accuracy of the vortex flowmeter within the range of the straight pipe section.

、参数提供不对 3 , the parameter is provided incorrectly

      Before purchasing the vortex flowmeter, the user must provide the parameter medium, flow rate, caliber, pressure, and temperature close to the on-site use values. Errors in the parameters cause the meter to calculate errors. The frequency is greater than the calculated full screen. The instructions fluctuate widely and cannot be read. Exact value.